- Germany
Non-contact Measurement
  RayDex hr
is anagolous in design to the RayDex cr, allows however though the lager rotor diameter even higher repeatabilities.

This version allows axial penetration depths (e.g. into a cylinder) of nearly 150 mm.

Applications of all RayDex Sensors:

  • Within measurement stations for individual objects in production and / or in a laboratory
  • Attached to a robot, to conduct interior measurements for e.g.
  • Mounted to a Coordinate Measurement Machine (CMM), once accuracy and speed are crucial factors
Specification  
Focus: RayDex hr
Spot Size 5 μm
Max. acquisition rate 4 kHz
Repeatability static* 0.1 μm
Repeatability dynamic1,* 0.2 μm
Measurement accuracy* <0.5μm
Max. speed of refocusing 200 μm / ms
Rotation:  
Repeatability in diameter (static)2* 0.1 μm
Repeatability in angular position* 0.0005 °
Max. rotation frequency 2.5 Hz
Axial directon:  
Axial range -
Repeatability in diameter (dynam.)3* 0.2 μm
Repeatability of an axial position4* < 1 μm
Max. axial speed 150 mm/s
* Accuracy and repeatability under the following constraints:
• Tolerance interval: 3 sigma
• 25 repetitions, acquisition rate: 250 Hz
• Vibration damping (e.g. granite plate)
• Fluctuations in temperature: < 0.5 K/h
• Variations in pressure: < 0.01 bar/h, fine-filtered, 30l/min

1 Starting position 2 mm out of focus, refocussing at maximum speed,
settling time of 50 ms.

2 Calibrated ring of 12 mm diameter; measuring rod of 1 pitch length and 6 mm diameter; rotational frequency is 3 Hz; axial position is static.

3 Calibrated ring of 12 mm diameter; measuring rod of 1 pitch length
and 6 mm diameter; rotational frequency is 3 Hz; axial position is
dynamic as described under 5.

4 Starting position 20 mm out of target position, axial motion with
max. speed; settling time of 50 ms; Readout from axial position
to an accuracy of < 0.5 μm.